STL2201 Hipot Tester (AC/DC/IR)

STL2201 series AC/DC withstanding voltage & insulation tester is a kind of Hipot Tester. Due to simple and compact structure, mature technique, brand new structure and operating interface, the operation becomes more convenient, and more practical functions are included as well. STL2201 series can be widely applied in transformer, device, component especially for winding safety inspection.

Key Features

240×64 Dot-matrix graphic LCD display

Rapid discharge and arc detection functions

Human body protection function

Features a built-in 8-channel matrix scanner for convenient use.

Voltage ramp-up time, test duration, and voltage ramp-down time can be set arbitrarily. The waiting time for current judgment during DC withstand voltage testing can be adjusted to accommodate loads with different characteristics.

50 sets, 100 steps per set, 500 steps in total.

Current baseline reset function

All-new operating interface and user-friendly panel design.

The instrument's operating program can be automatically upgraded via RS232C.

Module

STL2201

STL2201S ( Eight-way matrix scanner )

STL2201B

 

Output Voltage

AC: 0.05kV – 5.00kV ± ( 1.0% of reading+10V ) ( 50, 60Hz optional )

DC: 0.05kV – 6.00kV ± ( 1.0% of reading+10V )

Current Test Range

STL2201/STL2201S

AC: 0.001mA – 30mA

DC: 0.1 μA – 10mA

STL2201B

AC: 0.001mA – 20mA

DC: 0.1 μA – 5mA

Test accuracy:±(1%Reading + 5 characters)

Discharge function: Automatic discharge after the test (DCW)

Output Voltage

0.05kV – 1.00kV ±(1.0% of reading + 2 V)

Resistance Measurement Range

0.1MΩ – 50GΩ,(Current range: 10 nA–10 mA)

Resistance Measurement Accuracy

≥ 500V

1MΩ-1GΩ        ±(5% of reading + 5 digits)

1GΩ-10GΩ      ±(10% of reading + 5 digits)

10GΩ-50GΩ    ±(15% of reading + 5 digits)  (STL2201S: N/A)

< 500V

0.1 MΩ – 1 GΩ    ±(10% of reading + 5 digits)

 

Discharge Function

Automatic discharge after the test

Measurement Range

AC:

1mA – 15mA

DC:

STL2201/STL2201S: 1mA – 10mA

STL2201B: 1mA – 5mA

Voltage Rise Time

0.1s – 999s

Voltage Dip Duration

0.1s – 999s

Voltage Wait Time

0.1s – 99.9s (DC only)

Interface

RS232,USB,HANDLER,REMOTE I/O ,SCAN,GPIB (Optional)

Main Machine

Module No. Item Description
STL2201 Hipot Tester AC/DC/IR, including withstand voltage test cable
STL2201S Hipot Tester AC/DC/IR, including withstand voltage test cable
STL2201B Hipot Tester AC/DC/IR, including withstand voltage test cable