E-mail: sales@salukitec.com    Tel: +886. 909 602 109
Current Home > Vector Network Analyzer
Comparison Between S3601A Vector Network Analyzer and Keysight E5063A

This table will show the comparison between Saluki S3601A vector network analyer and Keysight E5063A.

Table 1 Product comparison

Specifications
Saluki S3601A
Keysight E5063A-235
Frequency Range100kHz - 3GHz100kHz - 3GHz
Frequency Resolution1 Hz1 Hz
Frequency Accuracy±5×10-6 (23°C±3°C)NA
Output Power-45 to +10dBm
-20 to 0dBm
System Dynamic RangeFrequency rangeIF=10HzIF=3kHzFrequency rangeIF=10HzIF=3kHz
100kHz - 1MHz9060100kHz - 300kHz8863
1MHz - 10MHz11080300kHz - 8.5MHz9368
10MHz - 3GHz125958.5MHz - 3GHz11792
Noise Floor
100kHz - 10MHz-90dBm
100kHz - 8.5MHz-103
10MHz - 3GHz-115dBm
8.5MHz - 3GHz-127
Harmonic< -18dBcNA
Effective Directionality
100kHz - 1MHz49100kHz - 1MHz49
1MHz - 3GHz46
1MHz - 3GHz46
Effective Source Match100kHz - 1MHz44100kHz - 1MHz41
1MHz - 3GHz401MHz - 3GHz40
Effective Load Match
100kHz - 1MHz49100kHz - 1MHz47
1MHz - 3GHz46
1MHz - 3GHz46
Reflection Tracking
100kHz - 1MHz±0.03100kHz - 1MHz±0.011
1MHz - 3GHz±0.021MHz - 3GHz±0.021
Transmission Tracking
100kHz - 1MHz±0.03100kHz - 1MHz±0.082
1MHz - 3GHz±0.021MHz - 3GHz±0.037

Magnitude Trace Noise

(Frequency=1GHz, IF=3kHz)

0.001 dB rms

Magnitude Trace Noise

(Frequency=1GHz, IF=70kHz)

0.005 dB rms

Phase Trace Noise

(Frequency=1GHz, IF=3kHz)

0.01 dB rms

Magnitude Trace Noise

(Frequency=1GHz, IF=70kHz)

0.009 dB rms
Analysis DomainFrequency/Time (standard)Frequency/Time (option)
IF Bandwidth1Hz - 5MHz (stepping by 1,2,3,5,7)
NA
Sweep Point1 -16001
NA
Port Connector Type
N type (Female), 50Ω impedanceN type (Female), 50Ω impedance
Measurement of Ports2 port (standard)2 port (standard)
Peripheral InterfaceUSB, GPIB, VGA, LANUSB, GPIB, VGA, LAN
Dimension (W*H*D)435*233*348mm432*235*277mm
Max. Power Consumption150W120W
Max. Weight16kg11kg